李瑞芳,黄存影,张文敏,李庆朝,刘希霖.色谱柱长度对二叔丁基过氧化二异丙基苯含量测定结果的影响[J].轮胎工业,2025,(7):0444-0447 |
 二维码(扫一下试试看!) |
色谱柱长度对二叔丁基过氧化二异丙基苯含量测定结果的影响 |
Influence of Chromatographic Column Length on Determination Results of Di Tert Butyl Peroxide Isopropylbenzene Content |
投稿时间:2024-03-04 修订日期:2024-03-04 |
DOI:10.12135/j.issn.1006-8171.2025.07.0444 |
中文关键词: 气相色谱法 BIBP 色谱柱长度 |
英文关键词: gas chromatography method BIBP length of column |
基金项目: |
中图分类号: |
|
摘要点击次数: 10 |
全文下载次数: 1 |
中文摘要: |
采用气相色谱法研究HP-5色谱柱长度对二叔丁基过氧化二异丙基苯(BIBP)的含量测定结果的影响。结果表明:当HP-5色谱柱长度为10 m时,BIBP峰型尖锐且对称,间位和对位BIBP同分异构体分离效果较好且检测结果的相对标准偏差均小于0.1%,试验结果的稳定性较好。 |
英文摘要: |
The effect of HP-5 chromatographic column length on the determination of di tert butyl peroxide isopropylbenzene(BIBP) content was studied by using gas chromatography.The results showed that when the HP-5 chromatographic column length was 10 m,the BIBP peak was sharp and symmetrica,and the separation of meta-and para-BIBP isomers were good,the relative standard deviations of the detection results were all less than 0.1%,indicating good stability of the experimental results. |
|